| Model Number |
8350322 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare professional (hcp) via a manufacturer representative regarding spinal devices used in spinal fusion.It was reported that the set screw was idling when the final was performed without the counter.Afterwards, the final was performed, but it seemed like the set screw was diagonal, so the set screw was replaced and the final was performed again, but it was idling again.The same procedure was repeated twice, and 2 screws were also replaced.It was also reported that, when removing the set screw, it was harder than usual and the 2 obturators were broken, so it would be returned.There was a delay of greater than 60 minutes in the overall procedure.Preoperative diagnosis of this surgery was scoliosis.Level at which implant was performed- t11-l3.There were no device fragments left in the patient body and no symptoms or complications were reported in this event.Additional information was received from the manufacturer representative that the second screw could not be tightened and replaced, setscrew idled and replaced, and the obturators were broken.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis: product# 8350322, lot# pr12m004 visually confirmed part of the tip has been broken off, consistent with interface during usage.Inspection of the material hardness confirmed conformance to print specification.Microscopic examination of the fracture surface identified a fairly flat fracture surface and circular material displacement.The above findings are consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|