• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W,WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W,WO CONNECTOR) Back to Search Results
Catalog Number 100/860/075CZ
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2020
Event Type  malfunction  
Manufacturer Narrative
Several device history records were reviewed with different lot numbers.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.A product sample and video were received for evaluation.Visual and functional testing were performed.Under visual inspection the sample appeared to be in good condition.On the video it is shown that it is not possible to inflate tracheostomy tube cuff.The complaint is confirmed.During manufacturing process the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received sample.It was observed that it is not even possible to inflate cuff because it leaks so badly from app.1mm long tear in cuff.Due to fact that cuff leak was found during use (not during pretesting as required by ifu) it is the most probable that reported failure occurred due to contact with sharp edge which is in conflict with instruction for use(ifu).Pma/510k is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event Description
It was reported that the material is defective in the balloon.The issue was identified immediately when the cannula was placed on the patient.There was no patient injury and the device was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W,WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15410935
MDR Text Key306015109
Report Number3012307300-2022-18949
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number100/860/075CZ
Device Lot Number3839070
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2022
Was the Report Sent to FDA? No
Date Manufacturer Received07/08/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-