SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number ZT19GN45ZGZ225N
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Operator of device is unknown.No information has been provided to date.Initial reporter also sent report to fda is unknown.No information has been provided to date.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product photo was received for evaluation.From review of the image, it was confirmed the wrong length was printed on the device.Unable to determine the root cause of the reported issue.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that product was labeled as 45mm, but actual length is 54mm.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15411020
Report Number3012307300-2022-18954
Device Sequence Number17892241
Product Code JOH
UDI-Device Identifier15019517200409
UDI-Public15019517200409
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date05/12/2025
Device Model NumberZT19GN45ZGZ225N
Device Catalogue NumberZT19GN45ZGZ225N
Device Lot NumberBS006690
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1