MAUDE Adverse Event Report: MANNKIND CORPORATION - V-GO V-GO DISPOSABLE INSULIN DELIVERY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MANNKIND CORPORATION - V-GO V-GO DISPOSABLE INSULIN DELIVERY DEVICE Back to Search Results
Model Number V-GO 40
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient reported the device released before 24-hours.Patient does not remember how long he wore it before it released.The patient reported that product samples are available for return to the manufacturer for evaluation.However; to date, no product sample has been received.
 
Additional Manufacturer Narrative
Device evaluation: one device was received and evaluated for the complaint regarding the needle button released event.Device #052408-a was received and needle mechanism function was tested and was verified to have operated as intended.The complaint could not be confirmed for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-GO DISPOSABLE INSULIN DELIVERY DEVICE
Common Device Name
V-GO
Manufacturer (Section D)
MANNKIND CORPORATION - V-GO
293 boston post road w
#330
marlborough MA 01752
Manufacturer (Section G)
MANNKIND CORPORATION - V-GO
293 boston post road w
#330
marlborough MA 01752
Manufacturer Contact
jeffrey zajac
293 boston post road w
#330
marlborough, MA 01752
MDR Report Key15411271
Report Number1226572-2022-00133
Device Sequence Number6744763
Product Code LZG
UDI-Device Identifier00385609600012
UDI-Public(01)00385609600012(17)220917(10)FG419232
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/19/2022
Device Model NumberV-GO 40
Device Catalogue Number8560-9600-01
Device Lot NumberVG419231B
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/19/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-