|
|
| Model Number |
174233 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a single port laparoscopic cholecystectomy, before the device was used to the patient, a scratch w as observed on the plastic covering the articulation part in front of the grasping part of the tip when it was opened.When the tip was retracted into the shaft in order to use, it was torn from the scratch, so use was discontinued.A new product was used to complete the case and resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the sheath of the instrument was damaged.Functionally, the jaw rotation worked without difficulty.Release lever and jaw toggle functioned properly.Ratchet button functioned properly.Jaw articulated, opened and closed without difficulty.It was reported that the plastic coating on the shaft of the device was ripped, torn, or the two ends of the reload of the endo catch ring connected by plastic shrink wrap was damaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if excessive manipulation is applied to jaws or device when articulating.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.Consult the medical literature relative to complications, hazards and techniques, prior to performing endoscopic procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a single port laparoscopic cholecystectomy, before the device was used to the patient, a scratch w as observed on the plastic covering the articulation part in front of the grasping part of the tip when it was opened.When the tip was retracted into the shaft in order to use, the shrink wrap was damaged, so use was discontinued.A new product was used to complete the case and resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted the sheath of the instrument was damaged.Functionally, the jaw rotation worked without difficulty.Release lever and jaw toggle functioned properly.Ratchet button functioned properly.Jaw articulated, opened and closed without difficulty.It was reported that the plastic coating on the shaft of the device was ripped, torn, or the two ends of the reload of the endo catch ring connected by plastic shrink wrap was damaged.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if excessive manipulation is applied to jaws or device when articulating.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.Consult the medical literature relative to complications, hazards and techniques, prior to performing endoscopic procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|