Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that this was a posterior cervical fusion performed on (b)(6) 2022.The nurse assembled the navigation drill guide body in question, but the silver knob part was pushed down while the scale of the drill guide was not set to "a".So, the doctor pushed the silver knob in to make an adjustment, then instruments got stuck where it shouldn't have been.The navigation drill guide body could no longer be disassembled.There was no surgical delay.No fragments was reported.Patient status/ outcome? :stable no further information is available.This report is for one (1) adj drill guide body 2.8mm.This is report 2 of 2 for complaint (b)(4).
|