Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
H3.Product analysis: equipment used: video inspection system (m-81805), ruler (m-83360), camera (panasonic lumix dmc-zs5 as found condition: the axium prime coil was returned for analysis within a shipping box and within a sealed plastic biohazard pouch.Visual inspection/damage location details: the axium prime pusher was found bent at ~7.4cm, ~37.1cm, ~778cm, and ~128.3cm from the proximal end.No damages or irregularities were found with the shield coil.The axium prime implant coil was found stretched and damaged with the polypropylene filament broken.Conclusion: the customer¿s report of ¿poor visualization¿ could not typically be confirmed through device analysis and no images or videos were submitted for analysis.Possible causes of poor visualization are imaging equipment, issues, material in the path, patient placement or position of implant.The pusher was found bent and the coil was found damaged/stretched.Possible causes of pusher bend and coil stretch/damage are lack of hydration, user does not maintain continuous flush, tortuous anatomy, coil not retracted in a one-to-one motion with the implant during repositioning, pushwire rotation or user advances the coil against resistance.The stryker sl-10 micro catheter was found kinked.It is possible the damaged micro catheter contributed to the coil/pusher damaged.The sl-10 micro catheter has an inner diameter of 0.016¿, which is compatible for use with the axium prime coil.H6.Coding updated based on the analysis results.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|