BOSTON SCIENTIFIC CORPORATION DIREXION HI-FLO FATHOM-16 SYSTEM; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number 84591
Medical Device Problem Codes Entrapment of Device (1212); Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
Reportable based on device analysis completed on 19aug2022.It was reported that the device got stuck inside the base catheter and was stretched.A direxion hi-flo fathom-16 system was selected for use.During the procedure, it was noted that the device got stuck inside the base catheter and eventually got stretched.It took time for the device to return to its initial position.No patient complications were reported.However, device analysis revealed 2 shaft fractures located 53.5cm and 68cm from the hub.
 
Additional Manufacturer Narrative
Date of event is an estimated date based on the aware date as the exact event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.The device returned stuck inside a 5fr cook angiographic catheter.The devices shaft showed stretching and 2 fractures located 53.5cm and 68cm from the hub.It was attempted to separate the direxion from the angiographic catheter; however, the two devices could not be separated.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIREXION HI-FLO FATHOM-16 SYSTEM
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15412091
Report Number2124215-2022-33892
Device Sequence Number6998958
Product Code KRA
UDI-Device Identifier08714729839712
UDI-Public08714729839712
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K142259
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2024
Device Model Number84591
Device Catalogue Number84591
Device Lot Number0028721418
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-