|
It was reported by the provider that the patient's right ijv vein was accessed easily with the needle with excellent blood return.The micro-wire was advanced assuring bevel in the "up" position.The wire was then found to be coiled in the vein so was withdrawn.On withdrawal, the wire "jammed" in the needle.The needle was then backed out slowly as the wire would not move and the wire was dissected free from any surrounding tissue directly at the puncture.Despite all efforts the wire refused to budge.Eventually with time and effort the wire came loose and the outer "wrap" of wire was found to be frayed and separated from the wire core.When the wire was able to be withdrawn, it was noticed a portion of the wire was missing and retained in the patient.Ultrasound and flouro were used to confirm this, and the fragment appeared to be stuck in the soft tissue between the vein wall and the skin.The provider determined to not attempt to remove the fragment due to its location within the vein lumen and behind the clavicle, and the risk outweighed the benefit of removing.
|
|
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.The customer's reported complaint description of guidewire tip detached cannot be confirmed.No guidewire sample was returned for evaluation.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.(b)(4) was sent to guidewire supplier/manufacturer, heraeus medical for awareness of this event and dhr review of supplier lot number.Heraeus confirmed no internal non-conformance reports were found regarding batch in question and personnel were properly trained.This event cannot be confirmed to be a manufacturing issue.Reference (b)(4) for details.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: direction for use {dfu} is provided with this mini-stick device.The failure mode of device fracture and embolization is cautioned as potential adverse event.No sample was returned for evaluation so it cannot be determined if device was used in a manner inconsistent with labeling.Adverse events guidewire shearing, fracture or embolization.Operational instructions: prior to insertion, flush all components with saline or heparinized/saline.Gain percutaneous access with the 21 g (0.9 mm) vascular introducer needle.Advance the 0.018 inch (0.46 mm) guidewire through the needle.Caution: the guidewire should not be withdrawn through the introducer needle.If the guidewire must be withdrawn while still inside the needle, simultaneously remove both the needle and the wire as a unit to prevent the needle from damaging the guidewire.Withdraw the introducer needle, leaving the 0.018 inch (0.46 mm) guidewire in place.Advance the coaxial assembly over the 0.018 inch (0.46 mm) guidewire.Simultaneously remove the dilator and 0.018 inch (0.46 mm) guidewire, while holding the sheath portion of the assembly in place.Advance a 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) guidewire into the sheath.Remove the sheath, leaving the 0.035 inch (0.89 mm) or 0.038 inch (0.97 mm) wire.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
|