• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MAXFORCE TTS; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MAXFORCE TTS; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00589150
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce tts dilatation balloon was used in the esophagus during an esophageal dilatation procedure performed on (b)(6) 2022.During the procedure, it was noted that the balloon ruptured.The procedure was completed with another maxforce tts dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h2: block e1 has been updated based on additional information received september 11, 2022.Block h6: device code a0402 captures the reportable event of balloon burst.
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce tts dilatation balloon was used in the esophagus during an esophageal dilatation procedure performed on august 24, 2022.During the procedure, it was noted that the balloon ruptured.The procedure was completed with another maxforce tts dilatation balloon.There were no patient complications reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that a maxforce tts dilatation balloon was used in the esophagus during an esophageal dilatation procedure performed on (b)(6) 2022.During the procedure, it was noted that the balloon ruptured.The procedure was completed with another maxforce tts dilatation balloon.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: device code a0402 captures the reportable event of balloon burst.Block h10: investigation result: the returned maxforce tts balloon was analyzed and a visual examination found that the balloon has no visual defects and was in good condition.The catheter of the device had no damages.Functional analysis was performed and the balloon was inflated without a problem.However, the balloon would not hold pressure due to a pinhole in the distal section of the balloon.Microscopic inspection found the balloon had a pinhole located approximately 30mm from the proximal bond.With all the available information, boston scientific concludes the reported event of balloon burst was not confirmed.It is possible that factors encountered during the procedure, such as the interaction with any surface during the procedure could create friction on the balloon and cause a pinhole.Therefore, the most probable root cause is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXFORCE TTS
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15412555
Report Number3005099803-2022-05138
Device Sequence Number5895107
Product Code KNQ
UDI-Device Identifier08714729099918
UDI-Public08714729099918
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K934697
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2024
Device Model NumberM00589150
Device Catalogue Number8915
Device Lot Number0028809174
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/24/2022
Supplement Date Received by Manufacturer09/11/2022
10/10/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date09/30/2022
11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-