• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION PRISMAFLEX SET (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION PRISMAFLEX SET (M); DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 106696
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that few minutes after starting therapy, one unit of prismaflex m60 set had a blood leak at the return line up near the return / venous clamp and just below that area.The amount of blood loss was estimated to be around 200ml.There was patient involvement however, no patient injury or medical intervention was reported.No additional information is available.
 
Additional Manufacturer Narrative
The actual sample was received for evaluation.A pressure test was performed, no leak was detected.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX SET (M)
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEYZIEU
av lionel terray, b.p. 126
7
meyzieu cedex rhone 69883
FR   69883
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15412830
Report Number8010182-2022-00262
Device Sequence Number12235865
Product Code KDI
UDI-Device Identifier07332414064549
UDI-Public(01)07332414064549
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number106696
Device Lot Number21K0069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PRISMAX
Patient SexUnknown
-
-