| Model Number |
SCPX127MH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: tissue expander deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient underwent an unspecified breast surgery with a mentor cpx4 plus, smooth, medium height 450cc tissue expander that deflated post implantation.The deflation was observed during explantation surgery on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
On 21-sep-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient had surgery and during it was discovered the expander was leaking (deflation).The product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned sample revealed that the tissue expander was found to have a tear on the anterior view, measuring 2.1 cm.Microscopic examination was performed on the edges of the tear, and parallel striations were found in an area of the tear, measuring approximately less than 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the device shell.The cause in the remaining area of the tear could not be identified.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.A second product was received (lot #9654345).No adverse events were reported for this concomitant (contralateral) device.Therefore, no further analysis is required.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|