It was reported that, during nailing procedure, three (3) 125 rad drill gde drop are extensively worn & unusable.These drops have seen extensive use over the past several years.They consistently miss the distal locking screws for short nails.Surgery was performed, after a significant delay, and with another surgical technique.Patient was not harmed as consequence of this problem.
|
|
The devices were not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, there were no clinical factors found which would have contributed to the event.According to report, the surgeon was able to complete the procedure by a change in the surgical technique after a significant delay.Since it was reported there was no patient harm because of this problem, no further clinical/medical assessment is warranted at this time.A review of complaint history of the previous 12 months revealed similar events for the listed devices, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.These devices are reusable instruments that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the devices to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual products involved, our investigation could not proceed.Should the devices or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|