Model Number 8696G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Events reported via 2210968-2022-07469, 2210968-2022-07470, and 2210968-2022-07471.
 
Event or Problem Description
It was reported that a patient underwent an unknown plastic surgery on (b)(6) 2022 and the suture was used.During surgery, the suture broke during use.There were no adverse patient consequences reported.Additional information has been requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 9/22/2022.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: please provide name and email of the initial reporter: no further information is available.Device return status? the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events reported via 2210968-2022-07468, 2210968-2022-07469, 2210968-2022-07470, and 2210968-2022-07471.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) date sent to the fda: 11/14/2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3 h3 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned product determined that it was received one paper lid and a needle-suture piece that pertain to the product code 8696g.During visual inspection of the returned sample, marks that appear to be caused by a surgical instrument were observed on the body needle.The suture was examined and the end of the suture was noted to be cut probably caused by a surgical instrument.In addition, body fluids were noted along the strand.As per the conditions of the returned sample, no functional test could be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 18IN 7-0 S/A P-1 PRM
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
982 road 183 km 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15414220
Report Number2210968-2022-07468
Device Sequence Number5640494
Product Code GAW
UDI-Device Identifier10705031019188
UDI-Public10705031019188
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8696G
Device Catalogue Number8696G
Device Lot NumberRPBBXH
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer09/15/2022
10/24/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date09/23/2022
11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-