Model Number VTICM5_13.2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Claim#: (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's right eye (od).The surgeon reports the patient is having difficulty reading/seeing; and was prescribed multiple medications with a planned follow up visit.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
B5 - the lens was exchanged for a different power lens.Reportedly, "patient tolerated the procedure well" without complications.H3 - device evaluation: lens was returned in a specimen cup with moisture.Visual inspection found the lens haptic torn.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key15414373
Report Number2023826-2022-03108
Device Sequence Number5653160
Product Code MTA
UDI-Device Identifier00841542117173
UDI-Public00841542117173
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model NumberVTICM5_13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-