| Model Number |
VTICM5_13.2 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's right eye (od).The surgeon reports the patient is having difficulty reading/seeing; and was prescribed multiple medications with a planned follow up visit.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - the lens was exchanged for a different power lens.Reportedly, "patient tolerated the procedure well" without complications.H3 - device evaluation: lens was returned in a specimen cup with moisture.Visual inspection found the lens haptic torn.H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|