Model Number 1013466-200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents and/or complaints reported from this lot.Based on the information provided, the investigation determined that the reported balloon rupture was likely due to circumstances of the procedure.It is likely that the rupture occurred due to interaction with the heavily calcified superficial femoral artery that was 80% stenosed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional armada 18 referenced are filed under separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified superficial femoral artery that was 80% stenosed.The 3x200mm armada 18 balloon was advanced without issue to the lesion and inflated once at 8 atmospheres when it ruptured.The device was removed.Then a 4x200mm armada 18 balloon was advanced without issue to the lesion and inflated once at 8 atmospheres when it ruptured.The device was removed.A rotablator was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15414421
Report Number2024168-2022-09656
Device Sequence Number14830873
Product Code LIT
UDI-Device Identifier08717648208027
UDI-Public08717648208027
Combination Product (Y/N)N
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number1013466-200
Device Catalogue Number1013466-200
Device Lot Number9092641
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/01/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/26/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-