As reported for this event by the customer, during a therapeutic laparoscopic sigma procedure the device probe tip broke off and fell into the patient.The broken piece was retrieved from the patient.The procedure was successfully and normally completed with new similar device.There was a slight prolongation of the procedure, which was not relevant.There is no harm or adverse impact to the patient or any other person.
|
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Please see the updates in sections d9, h3, h4, h6, and h10.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the probe was broken around the outer pipe.Additionally, the coating of the grasping section was partially peeled off; however, this defect is not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Although a definitive root cause cannot be identified, the following step-by-step scenario likely caused the event: 1) during the output activation in seal & cut mode, the distal end of the probe was slightly contacting a device with a thin tip, causing spark generation.2) a force was applied to the probe during spark generation.3) a force to grasp tissue or a force to activate the output in seal & cut mode was applied to the probe.This generated cracks on the probe.4) due to continuous use of the device, the cracks on the probe progressed.This led to breakage of the probe.The following information is included in the instructions for use (ifu): ¿do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.¿ ¿if the grasping section, metal-exposed area around it or the probe tip gets sticked tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ olympus will continue to monitor the performance of this device.
|