ALERE SAN DIEGO, INC. ALERE HCG COMBO CASSETTE (20/10MIU/ML); PREGNANCY TEST Back to Search Results
Model Number FHC-A202
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Customer unresponsive.Pending results of investigation.
 
Event or Problem Description
Unspecified date: customer reported several false positive results when testing with the alere hcg combo cassette.No adverse outcomes reported.The exact frequency or number of patients involved was not provided.Although requested, further information was not provided at intake.
 
Additional Manufacturer Narrative
Retained devices from the reported lot number were tested with hcg-negative clinical urine and hcg-negative clinical serum samples.The results were read at 3 and 4 minutes for devices with urine samples and 5 and 6 minutes for devices with serum samples.All devices yielded valid negative results.No false positive results were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the packet insert: very low levels of hcg (less than 50 miu/ml) are present in serum and urine specimen shortly after implantation, however, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning serum or urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in serum or urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (hama) in the specimen.Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain hama.Such specimens may cause false positive or false negative results.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALERE HCG COMBO CASSETTE (20/10MIU/ML)
Common Device Name
PREGNANCY TEST
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
monika burrell
9975 summers ridge road
san diego, CA 92121
8588052506
MDR Report Key15414645
Report Number2027969-2022-00077
Device Sequence Number20247179
Product Code JHI
UDI-Device Identifier10815845020123
UDI-Public(01)10815845020123(17)221231(10)HCG0122137
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K062361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model NumberFHC-A202
Device Lot NumberHCG0122137
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/02/2022
Supplement Date Received by Manufacturer10/13/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-