| Model Number |
PS TIBIAL INSERTS SZ 4, 13MM |
| Medical Device Problem Codes |
Unstable (1667); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526)
|
| Date of Event |
01/01/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had a initial left tka on (b)(6) 2010.In approximately 2016, the patient had a revision due to polymer wear and laxity.No other patient information/medical history provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.200-02-32 three peg patella 32mm 1769025.204-04-44 trapezoid tibial tray sz 4f/4t 1588592.204-34-04 fluted stem extension 40l x 14 mm 1662098.204-70-00 tibial stem ext.Screw 1712470.234-02-04 optetrak asy, ps cemented femoral, sz 4, 1628337.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Investigation results - the revision reported may have been the result of polyethylene wear and instability as stated in the operative notes.The extent and root cause of the reported polyethylene wear cannot be determined as the devices were not available for evaluation, and pre-revision radiographs and images of the explanted devices were not provided.These devices are used for treatments, not diagnosis.There is no other information available.
|
| |
|
Event or Problem Description
|
|
As reported, the patient had a initial left tka.The patient had a revision due to polymer wear and laxity.No other patient information/medical history provided.No device returned.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b3 (remove date - unknown), d6b (remove date - unknown), f10 (remove data from initial), g1, g2, h6: clinical code, medical device problem code, component code, investigation findings.D6b: patient has been revised.Date is unknown at this time.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|