• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 67PFS40
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were white and brown stains on the cuff after sterilization.Before sterilization, the tube was washed with mild soap and water.The tube was unusable and another one was used.No patient injury was reported.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspection were completed and no issues were noted during manufacture.The sample was received in used conditions, with its certificate of safe handling and without its original packaging inside a ziploc bag.Visual inspection was performed at 12" under normal lighting to received unit, in order to detect any damage on the cuff or airline.Results: during the visual inspection, we observed yellow stains in the cuff.Contamination not well measured for this occurrence the mitigation to detect the reported complaint.Components are 100% inspected by production personnel before and after every operation to detect any contamination.If during the inspection, they detect contamination in the parts, it is measured to ensure that meet with the specification of finished goods.Components are 100 percent inspected by production personnel before and after every operation to detect any contamination in the units.Quality inspectors takes a representative sample of the lot and performs a visual inspection in order to detect any contamination on the units according to inspection for finished goods.The most probable root cause is that damaged occurred after the product left the facility.No corrective actions were taken.No information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15414726
Report Number3012307300-2022-19019
Device Sequence Number5512380
Product Code JOH
UDI-Device Identifier10351688518620
UDI-Public10351688518620
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/19/2024
Device Model Number67PFS40
Device Catalogue Number67PFS40
Device Lot Number3735240
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-