Model Number 72404291
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the hospital with an inflatable penile prosthesis (ipp) that was not functioning properly.At a later date, the cylinder and pump were explanted and replaced with a new cylinder and pump.Upon explant, a crack was found in the cylinder connecting tube.No patient complications were reported.
 
Event or Problem Description
It was reported that the patient presented to the hospital with an inflatable penile prosthesis (ipp) that was not functioning properly.At a later date, the cylinder and pump were explanted and replaced with a new cylinder and pump.Upon explant, a crack was found in the cylinder connecting tube.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt of the inflatable penile prosthesis (ipp) at our quality assurance laboratory, the returned components underwent a thorough analysis.Visual examination of one of the cylinders identified multiple leaks consistent with sharp instrument damage on one cylinder.The other cylinder presented a leak in the kink resistant tubing (krt) that was also consistent with sharp instrument damage.The pump krt had fatigued and was worn down to the filament, but the pump did not present any leaks.The pump did not pass the functional testing that was performed.Based on the information available, the pump did not operate within product specifications and could affect the functionality of the device; therefore, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Additional Manufacturer Narrative
Upon receipt of this inflatable penile prosthesis (ipp) at our quality assurance laboratory, the returned components underwent a thorough analysis.Visual examination of one of the cylinders identified multiple leaks consistent with sharp instrument damage.The other cylinder had a leak in the kink resistant tubing (krt), also consistent with sharp instrument damage.The tubing had fatigued and was worn down to the filament, but the pump did not present any leaks and passed the functional testing that was performed.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Event or Problem Description
It was reported that the patient presented to the hospital with an inflatable penile prosthesis (ipp) that was not functioning properly.At a later date, the cylinder and pump were explanted and replaced with a new cylinder and pump.Upon explant, a crack was found in the cylinder connecting tube.No patient complications were reported.
 
Event or Problem Description
It was reported that the patient presented to the hospital with an inflatable penile prosthesis (ipp) that was not functioning properly.At a later date, the cylinder and pump were explanted and replaced with a new cylinder and pump.Upon explant, a crack was found in the cylinder connecting tube.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt of the inflatable penile prosthesis (ipp) at our quality assurance laboratory, the returned components underwent a thorough analysis.Visual examination of one of the cylinders identified multiple leaks consistent with sharp instrument damage.The other cylinder presented a leak in the kink resistant tubing (krt) that was also consistent with sharp instrument damage.The pump krt had fatigued and was worn down to the filament, but the pump did not present any leaks and passed the functional testing that was performed.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15414754
Report Number2124215-2022-35807
Device Sequence Number6619314
Product Code FHW
UDI-Device Identifier00878953003931
UDI-Public00878953003931
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 12/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/26/2021
Device Model Number72404291
Device Catalogue Number72404291
Device Lot Number119417003
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/29/2022
Supplement Date Received by Manufacturer11/01/2022
11/10/2022
12/08/2022
Initial Report FDA Received Date09/13/2022
Supplement Report FDA Received Date11/11/2022
11/14/2022
12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/11/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PUMP LOT #146390001; PUMP LOT #146390001; PUMP LOT #146390001; PUMP LOT #146390001; RESERVOIR LOT #144010003; RESERVOIR LOT #144010003; RESERVOIR LOT #144010003; RESERVOIR LOT #144010003
Outcome Attributed to Adverse Event Hospitalization; Hospitalization; Required Intervention; Required Intervention;
Patient Age61 YR
Patient SexMale
-
-