| Model Number |
67SP040 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that cuff would not inflate and the water would not go down the cuff.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No lot number was provided; therefore, device history record review could not be completed.A product sample was received for evaluation.Visual and functional testing were performed.No visible damage detected.During functional test of product, cuff was observed to inflate asymmetrically outside of specifications.The root cause of the reported issue was found to be a deficiency during manufacturing process.This remediation mdr was generated under protocol (b)(4)., as a result of warning letter cms# (b)(4).
|
| |
|