• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 750170
Medical Device Problem Code Volume Accuracy Problem (1675)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One sample was received without its original packaging in used conditions.No damage was detected during visual inspection.When measuring the cuff, the percentage for the symmetry was 41.17%.This result is over 33.3% that is the minimum acceptable for air cuff.Based on the test, the unit is within spec.The complaint is not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Operator of device is unknown.Initial reporter also sent report to fda is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that device's balloon was inflating asymmetrically.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA TRACHEOSTOMY TUBE, NEONATAL/PEDIATRIC
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15415070
Report Number3012307300-2022-19036
Device Sequence Number2556077
Product Code JOH
UDI-Device Identifier15021312006414
UDI-Public15021312006414
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K914088
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/28/2024
Device Model Number750170
Device Catalogue Number750170
Device Lot Number3791379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-