• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRASCORE 014; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRASCORE 014; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number US1415074
Medical Device Problem Codes Material Rupture (1546); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure in the right common femoral artery via a contralateral approach, the pta balloon allegedly ruptured at 4atm.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 03/2024).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the right common femoral artery via a contralateral approach, the pta balloon allegedly ruptured at 4atm.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: (expiry date: 03/2024).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device pending return.
 
Additional Manufacturer Narrative
H10: additional information was received and the file was reassessed for reportability and determined to be reportable as malfunction.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultrascore 014 pta dilatation catheter was received for evaluation.During visual evaluation, the strain relief was noted to be missing from the catheter.No kinks were noted to the catheter shaft and no anomalies to luers/y-body.During functional evaluation, an in-house presto inflation device was used to inflate the balloon and pinhole rupture was noted to the balloon.No other functional testing was performed.Therefore, the investigation is confirmed for the reported uncomplicated pinhole balloon rupture as a pinhole rupture was noted to the balloon.The investigation is also confirmed for the identified strain relief dislodgement as the strain relief was noted to be missing from the catheter.A definitive root cause for the reported uncomplicated pinhole balloon rupture and identified strain relief dislodgement could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date: 03/2024), g3, h2, h6 (device).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the right common femoral artery via a contralateral approach, the pta balloon allegedly ruptured at 4 atm.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE 014
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15415759
Report Number2020394-2022-00738
Device Sequence Number20450572
Product Code PNO
UDI-Device Identifier00801741114120
UDI-Public(01)00801741114120
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS1415074
Device Catalogue NumberUS1415074
Device Lot NumberGFGQ0995
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer09/15/2022
10/14/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/12/2022
10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-