| Model Number |
21790400 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On 13-sep-2022: based on the available information, the malfunction did lead to, or might led to, death or a serious injury.A device history register review have been completed, and no deviation or changes were found during manufacturing of the device.Further investigations needed.This is a late report.
|
| |
|
Event or Problem Description
|
|
On 13-sep-2022: it was reported that when hearing aids were charging overnight into the next morning the charger was melted.The incident happened on (b)(6) 2022, and was replaced the same day.
|
| |
|
Event or Problem Description
|
|
13-sep-2022: it was reported that when hearing aids were charging overnight into the next morning the charger was melted.The incident happened on 08-aug-2022, and was replaced the same day.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers ref# (b)(4).13-sep-2022: based on the available information, the malfunction did lead to, or might led to, death or a serious injury.A device history register review have been completed, and no deviation or changes were found during manufacturing of the device.Further investigations needed.This is a late report.11-oct-2022: added code ''analysis of production records'' in types of investigation in h6.Added ''manufacturers ref# (b)(4) to narrative.Unticked h7 as no remedial actions have been done.Changed h8 to ''initial use of device''.Pending investigation results.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers ref#(b)(4).(b)(6) 2022: based on the available information, the malfunction did lead to, or might led to, death or a serious injury.A device history register review have been completed, and no deviation or changes were found during manufacturing of the device.Further investigations needed.This is a late report.(b)(6) 2022: added code ''analysis of production records'' in types of investigation in h6.Added ''manufacturers ref# (b)(4)'' to narrative.Unticked h7 as no remedial actions have been done.Changed h8 to ''initial use of device''.Pending investigation results (b)(6) 2022: corrected date in g3.Pending investigation results (b)(6) 2022: technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.No investigations done on the actual device.Trended case investigation results: the returned apollo 5.3 charger is returned because the usb-c connector is broken down.The returned device has signs of heating around the usb-c connector, also the cable shows signs of heating.The charger is fully functional except of the connector.It can charge two hearing aids.The internal battery is working.The usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector.The clinical investigation concluded: a damaged power supply or a power supply of very poor quality could potentially lead to burn injuries and in rare conditions there is a small risk that a wrong charging or discharging or a battery thermal runaway can lead to high temperatures or fire accidents.The chargers have been safety tested according to applicable standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.There are no new clinical aspects (e.G., unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusion herein are sufficient.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register based on the information the case is still considered reportable as the event / incident did lead to, or might led to, death or a serious injury.No further information is expected.
|
| |
|
Event or Problem Description
|
|
(b)(6) 2022: no further information is expected.(b)(6) 2022: it was reported that when hearing aids were charging overnight into the next morning the charger was melted.The incident happened on (b)(6) 2022, and was replaced the same day.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturers ref#: (b)(4).13-sep-2022: based on the available information, the malfunction did lead to, or might led to, death or a serious injury.A device history register review have been completed, and no deviation or changes were found during manufacturing of the device.Further investigations needed.This is a late report.11-oct-2022: added code ''analysis of production records'' in types of investigation in h6.Added ''manufacturers ref#: (b)(4)" to narrative.Unticked h7 as no remedial actions have been done.Changed h8 to ''initial use of device''.Pending investigation results.09-nov-2022: corrected date in g3.Pending investigation results.
|
| |
|
Event or Problem Description
|
|
13-sep-2022: it was reported that when hearing aids were charging overnight into the next morning the charger was melted.The incident happened on (b)(6) 2022, and was replaced the same day.
|
| |
|