| Model Number |
D002 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.As a result, device replacement within 90 days or more frequent monitoring was recommended.This icd remains in-service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.As a result, device replacement within 90 days or more frequent monitoring was recommended.This icd remains in-service.No adverse patient effects were reported.Additional information was received.This patient died of unrelated causes.This icd remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis; therefore, a technical analysis could not be performed.
|
| |
|