| Model Number |
1516-40-708 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon tried to implant the poly insert but it would not fit even though we trialed through a range of motion with the exact same size trial insert.So the surgeon asked for me to open a new one since there seemed to be a malfunction with the original poly insert.There was a reported surgical delay of about 10 minutes.Doe: (b)(6) 2022.Affected side: right knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated that the word ¿malfunction¿ was used because it seemed like something was wrong with the insert since the surgeon could not seat it into the patient even though an exact same trial size insert was used which seated normally.The surgeon assumed that it was not sliding in the knee like the normal.No harm was done to the patient.This occurred during a primary surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (nc search) was performed for the finished device 151640708, lot m04j08, and no non-conformances / manufacturing irregularities were identified.
|
| |
|