Model Number SC710 VICTORY 10 4 WHEEL
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 09/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Provider alleges consumer alleges his scooter was on his car carrier to go to his clinic.Once there, the scooter allegedly rolled off the carrier and allegedly ran over his toe.Consumer turned allegedly trying to catch his scooter and allegedly injured his hip.
 
Additional Manufacturer Narrative
The device has not been made available for evaluation.Should further information of the device become available, a follow-up report will be issued.
 
Event or Problem Description
Provider alleges consumer alleges his scooter was on his car carrier to go to his clinic.Once there, the scooter allegedly rolled off the carrier and allegedly ran over his toe.Consumer turned allegedly trying to catch his scooter and allegedly injured his hip.
 
Additional Manufacturer Narrative
The unit freewheeled while in drive upon return.This was due to the brake handle being severely bent to the point it's throw was impeded by the rear motor cover.(customer misuse) when the cover was removed, the handle was able to reach the full engaged position and the unit functioned as designed.The unit's regenerative braking system functioned as design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
MOTORIZED THREE WHEEL
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
Manufacturer (Section G)
N/A
n/a
n/a
n/a
Manufacturer Contact
kelly livingston
401 york ave
duryea, PA 18642
8008008586
MDR Report Key15417558
Report Number2530130-2022-00076
Device Sequence Number12443021
Product Code INO
UDI-Device Identifier00606509300346
UDI-Public00606509300346
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K071949
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberSC710 VICTORY 10 4 WHEEL
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/25/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-