• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HD SYS. W/O CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HD SYS. W/O CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 191128
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
A user facility biomedical technician (biomed) reported that blood loss occurred during hemodialysis (hd) treatment a fresenius 2008t machine.A blood clot was identified in the tubing of the bloodlines (brand unknown) and the biomed stated that blood loss was also identified.The machine initially alarmed with an arterial pressure alarm message and the patient was transferred to a different machine to complete treatment.It was reported that there were no adverse reactions experienced by the patient.Additional information was requested, however a response was not received.The patient's estimated blood loss (ebl) was not reported.No parts were returned to the manufacturer for a physical evaluation.
 
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.A causal relationship between the objective evidence and the alleged event could not be determined; the alleged event is unconfirmed.
 
Event or Problem Description
A user facility biomedical technician (biomed) reported that blood loss occurred during hemodialysis (hd) treatment a fresenius 2008t machine.A blood clot was identified in the tubing of the bloodlines (brand unknown) and the biomed stated that blood loss was also identified.The machine initially alarmed with an arterial pressure alarm message and the patient was transferred to a different machine to complete treatment.It was reported that there were no adverse reactions experienced by the patient.Additional information was requested, however a response was not received.The patient's estimated blood loss (ebl) was not reported.No parts were returned to the manufacturer for a physical evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. W/O CDX BLUE STAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jessica trujillo
920 winter st
waltham, MA 02451
6174175172
MDR Report Key15417613
Report Number0002937457-2022-01505
Device Sequence Number15079882
Product Code KDI
UDI-Device Identifier00840861102105
UDI-Public00840861102105
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173972
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number191128
Device Catalogue Number191128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-