• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. BIVONA; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 60SP040
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No lot number was provided; therefore, device history record (dhr) review could not be completed.A product sample and three pictures were received for evaluation.In the pictures, a big split was observed in the flange.The sample was used and without its original packaging.Visual and functional testing were performed.Visual inspection was performed under normal conditions of illumination in order to detect any damage on the parts.During the visual inspection, we observed a big split in the flange.The complaint is confirmed.A definitive root cause of the issue could not be determined.However, this product failure will be continuously be monitored for threshold or escalation.Operator of device is unknown.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that immediately after opening the package, the customer noticed the flange was torn.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
2-3-2 kamiochiai, chuoh-ku,
minneapolis, MN 55442
MDR Report Key15417629
Report Number3012307300-2022-19106
Device Sequence Number2556160
Product Code JOH
UDI-Device Identifier15021312000382
UDI-Public15021312000382
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K912469
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number60SP040
Device Catalogue Number60SP040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2022
Initial Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-