• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® LUBRICATH® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® LUBRICATH® FOLEY CATHETER Back to Search Results
Model Number 365716
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
 
Event or Problem Description
It was reported that there was a detachment of device or device component.Per additional information received via email on 30aug2022, stated that the balloon was lodged in patient¿s urethra and was unable to deflate balloon.The user consulted with urology and was instructed to cut the catheter to see if balloon would deflate.Still the balloon would not deflate.Patient had to have needle aspiration to deflate the balloon.
 
Event or Problem Description
It was reported that there was a detachment of a device or device component.Per additional information received via email on 30aug2022, stated that the balloon was lodged in patient¿s urethra and was unable to deflate the balloon.The user consulted with urology and was instructed to cut the catheter to see if the balloon would deflate.Still, the balloon would not deflate.The patient had to have needle aspiration to deflate the balloon.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.It is unknown whether the device had met relevant specifications.The product was used for urological care.It was unknown whether the product had caused the reported failure.Potential root cause for this failure mode could be user related (example: over aspirated, incorrect syringe/collapse lumen/sac close eye/valve damage).The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter maintenance 1) secure the foley catheter.Use the statlock® foley stabilization device if provided.2) maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions.3) maintain unobstructed urine flow and keep the catheter and collection tube free from kinking.4) keep the collection bag below the level of the bladder or hips at all times.5) empty the collection bag regularly using a separate, clean collection container for each.Foley catheter removal 1) to deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required tomake the syringe ¿stick¿ in the valve 2) allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently 3) use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation 4) if the balloon will not deflate and if permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol 5) should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15417720
Report Number1018233-2022-07094
Device Sequence Number19469146
Product Code EZC
UDI-Device Identifier00801741077746
UDI-Public(01)00801741077746
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/28/2024
Device Model Number365716
Device Catalogue Number365716
Device Lot NumberMYDS5542
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date09/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-