Model Number SA60AT
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, the iol had a scratch directly in the center of the lens.The scratch was observed after insertion when the lens unfolded.The iol was removed during the initial procedure.Additional information has been received and stated that, the lens was replaced with another lens during initial procedure.There is no patient harm reported.
 
Additional Manufacturer Narrative
The product was returned for analysis.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case.Solution was dried on the lens.One haptic was broken in the gusset area (not returned).The optic was cut into pieces, typical of insertion and removal.The cut optic pieces had multiple scratches on the anterior and posterior sides of the lens.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Information was provided in the file that indicated the use of a non-qualified lens/cartridge/viscoelastic combination.The company lens model is only qualified for use in the company model ii and iii cartridges.Company ovd (ophthalmic viscosurgical device) is the only qualified viscoelastic with the qualified lens/cartridge combination.The handpiece indicated is qualified when used with the qualified lens/cartridge combinations.The reported lens damage was observed.The root cause for the reported complaint appears to be related to failure to follow the ifu (instructions for use).The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.Company recommends using the qualified company iol delivery system or any other company qualified combination.There are no other complaints in this lot.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15417828
Report Number1119421-2022-01952
Device Sequence Number6618895
Product Code HQL
UDI-Device Identifier00380652555784
UDI-Public00380652555784
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.190
Device Lot Number15374021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient Age63 YR
Patient SexMale
-
-