C.R. BARD, INC. (COVINGTON) -1018233 UROFORCE® BALLOON DILATION CATHETER Back to Search Results
Model Number 887610
Medical Device Problem Codes Burst Container or Vessel (1074); Material Puncture/Hole (1504)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that there was a leak in a dilation catheter balloon.
 
Additional Manufacturer Narrative
The reported event is confirmed, cause unknown.The used balloon dilation catheter was returned in the opened original packaging.Visual evaluation noted some inflation medium present in the balloon along with some crystal like sediment.The balloon itself felt "crisp" to touch and flakes were noted to be falling off/out of the balloon.Although a specific cause cannot be determined, a potential root cause for this event could be, "poor balloon design".No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warnings: do not exceed the recommended maximum balloon inflation pressure that is stated on the label.Extreme caution should be used when considering dilation of irregular, non-compliant tissue or in the presence of certain calculi." "precautions: this device should be used only by a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract." "catheter balloon inflation: 1.Confirm proper placement of the balloon within the urinary tract.2.Inflate the balloon with standard inflation medium (e.G., diatrizoate meglumine injection u.S.P.60% diluted as appropriate) using a 10cc or larger syringe or inflation device.Note: do not use air or any gaseous substance as a balloon inflation medium.Note: the use of an inflation device to monitor balloon pressure is strongly recommended to determine that adequate pressure is applied and that maximum limits of the balloon are not exceeded.Caution: if loss of pressure in the balloon occurs or the balloon ruptures, immediately stop the procedure, deflate the balloon and remove carefully.Do not attempt to reinflate the balloon.If, upon inspection of the balloon, it is noted that pieces of the balloon are missing, check for and endoscopically retrieve the fragments when possible." "caution: if resistance is felt when either removing the guidewire from the catheter or the catheter from the endoscope, stop and consider removing them as a single unit to prevent damage to the product or tissue trauma.Applying excessive force to the catheter can result in tip breakage or balloon separation." correction: d,f,g,h h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event or Problem Description
It was reported that there was a leak in a dilation catheter balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UROFORCE® BALLOON DILATION CATHETER
Common Device Name
BALLOON DILATION CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15417843
Report Number1018233-2022-07098
Device Sequence Number6618905
Product Code EZN
UDI-Device Identifier10801741053020
UDI-Public(01)10801741053020
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K993840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/01/2025
Device Model Number887610
Device Catalogue Number887610
Device Lot NumberGFGS1697
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/01/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-