|
Upon receipt, the lead under complaint was found cut 53 cm proximal to the lead tip.The distal fragment was received for analysis.The proximal fragment was not returned.Further cuts into the insulation were found between 34 cm and 30 cm proximal to the lead tip.The inspection revealed that the insulation was, apart from the present cuts, free of breaches.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported clinical observations.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|