| Model Number |
FT19AN60NSF077N |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was uneven inflation found in multiple cuffs.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
No information has been provided to date.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.There were no issues identified with cuff inflation; all cuffs inflated.Root cause cannot be determined since the complaint was not confirmed due to the fact the sample was tested and successfully passed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.
|
| |
|
Search Alerts/Recalls
|