• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40295
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 08/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that guidewire tip detachment occurred.The stenosed target lesion was located in the pulmonary artery.A.018/260cm platinum plus guidewire was selected for use.However, during the removal of cardiomems implant sheath, the proximal tip of the wire broke off and was left in the pulmonary artery.The physician decided not to retrieve the detached wire.The procedure was completed with no patient complications reported and the patient was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: the device was returned for analysis.Unit returned with its original pouch, overall visual revision did not identify failures or evidence that could be lost due to decontamination process.The wire returned was inspected and it is possible to observe the core wire was detached at the distal tip.The outer dimension dimensions were found within specification.
 
Event or Problem Description
It was reported that guidewire tip detachment occurred.The stenosed target lesion was located in the pulmonary artery.A.018/260cm platinum plus guidewire was selected for use.However, during the removal of cardiomems implant sheath, the proximal tip of the wire broke off and was left in the pulmonary artery.The physician decided not to retrieve the detached wire.The procedure was completed with no patient complications reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM PLUS
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15418649
Report Number2124215-2022-32326
Device Sequence Number12416947
Product Code DQX
UDI-Device Identifier08714729326212
UDI-Public08714729326212
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K945379
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/01/2025
Device Model Number40295
Device Catalogue Number40295
Device Lot Number0029321276
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/22/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-