STRYKER NEUROVASCULAR CORK EXCELSIOR XT-27 STD PRE-SHAPED 150CM; CATHETER, PERCUTANEOUS Back to Search Results
Model Number M003XT2750910
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The device was received.The reported lot number was confirmed from the hub and from the packaging label.During visual inspection, the catheter shaft was found to be severely kinked/bent.The tip of the catheter was found to be damaged and broken/fractured 155 cm form the proximal end.The radio opaque (ro) marker was detached/fractured.The hub was found to be intact.Catheter difficulty advancing guidewire functional test was unable to perform due to condition of the device.Catheter tip damaged functional test was confirmed during visual inspection.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was confirmed based on analysis.The device failed to meet specification when received for complaint investigation based on the analyzed anomalies noted to the device.Additional information provided by the customer indicated that there was no damage noted to the packaging prior to opening and the device was confirmed to be in good condition during preparation/prior to use on the patient.The damage noted to the device would be indicative of the as reported defect.It was seen that the microcatheter was severely damaged, and part of the distal shaft and the ro marker was detached.The as reported events of catheter difficulty advancing guidewire and catheter tip damaged as well as the as analysed events of ro marker(s) detached/separated/not visible under fluoroscopy, catheter tip broken/fractured during use, catheter shaft kinked/bent, catheter tip damaged will be assigned procedural factors as these defects appear to be associated with a product that met stryker and design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural or anatomical factors during use.
 
Event or Problem Description
Analysis of the returned device revealed that subject catheter tip was broken/fractured during use and radio opaque (ro) markers(s) detached/separated/not visible under fluoroscopy.There was no clinical consequence to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCELSIOR XT-27 STD PRE-SHAPED 150CM
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15418967
Report Number3008881809-2022-00449
Device Sequence Number6581546
Product Code DQY
UDI-Device Identifier04546540698155
UDI-Public04546540698155
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K113778
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2024
Device Model NumberM003XT2750910
Device Catalogue NumberM003XT2750910
Device Lot Number23249404
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/24/2022
Initial Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SYNCHRO2 GUIDEWIRE (STRYKER)
Patient Age55 YR
Patient SexMale
Patient Weight60 KG
Patient RaceAsian
-
-