| Model Number |
TIBIAL INSERT FB SZ 2 RT 6MM |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Extraskeletal ossification (1428); Failure of Implant (1924); Pain (1994); Joint Impingement (4834)
|
| Date of Event |
03/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had an initial ankle replacement on (b)(6) 2018.The patient presented with medial symptoms, mechanical in nature, but related to the superficial peroneal nerve.Radiographs show heterotopic ossification of medial gutter, causing impingement.The patient was revised on (b)(6) 2022, and the tibial insert and locking clip were replaced.This event report was received through clinical data collection activities.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.Medical products: 350-02-02 - talar implant sz 2 rt 5148428, 350-10-01 - ankle sz 1 locking clip 5188250, 350-12-01 - tibial plate fb sz 1 rt 5238976, 351-91-03 - recipsawblade 8x50x1mm 239091.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of surgical revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.H7: z-0024-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b5, f10 (remove data from initial), g1, h7 (uncheck), h6: clinical code, medical device problem code, component code, type of investigation, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that a female patient had a revision surgery for impingement.Indication for initial implant surgery was post traumatic arthritis.Patient had medial symptoms mechanical in nature but related to the superficial peroneal nerve.Radiographs show heterotopic ossification of medial gutter, causing impingement.It is indicated that this is unlikely related to the device and possibly related to the procedure.Tibial insert and tibial locking clip were replaced.Outcome: continuing.No additional information available.
|
| |
|