• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT ORTHOPEDICS INC. PROFEMUR® TL STEM SZ 4; HIP COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT ORTHOPEDICS INC. PROFEMUR® TL STEM SZ 4; HIP COMPONENT Back to Search Results
Model Number PRTL0024
Medical Device Problem Code Corroded (1131)
Health Effect - Clinical Codes Pain (1994); Osteopenia/ Osteoporosis (2651); Metal Related Pathology (4530)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event will be updated once the investigation is complete.Trends will be evaluated.
 
Event or Problem Description
Allegedly, on (b)(6) 2012, patient had a wright medical artificial hip implanted in the right side of his body.On or about (b)(6) 2020 patient was told that he suffered a broken screw holding the dynasty cup, leading to loosening of the cup and pain in his right hip, and that it was advisable that he undergo surgery to remove the dynasty cup and screws and replace and refix the cup and fixation screws (it is not indicated which screw was broken since patient had 3 screws implanted.Also, it has not been indicated which cup and screws were used to replace the dynasty cup and screws).This event is captured under microport orthopedics incident group number: (b)(4).On or about (b)(6) 2022, patient suffered a right hip implant dislocation.By (b)(6) 2022, patient developed symptoms of, among others, low back pain, development of a cyst on his right thigh, in addition to suffering recent right hip dislocation.Patient's orthopedic surgeon after the (b)(6) 2022 dislocation, informed patient that in his medical judgment it was advisable to undergo a right hip revision surgery to remove the profemur® hip system, including the profemur® plus cocr modular neck and the profemur® tl stem, and replace it with a hip stem system of a different medical device manufacturer, smith & nephew.On (b)(6) 2022, patient underwent a revision surgery.In his (b)(6) 2022 revision surgery, operative report, surgeon noted failure of right hip implant and metallosis of right hip joint.During surgery on (b)(6) 2022 but just prior to actual removal of patient's wright medical hip implant, surgeon followed a path of accumulated fluid down the back of patient's right hip and observed grey necrotic tissue in the back of the hip.After removal of the femoral head and the modular neck, surgeon noted "a significant amount of corrosion around the modular portion of the neck that mates with the stem.It is all blackened metallosis.It made black streaks on my glove pulling this out." surgeon was able to extract the modular neck from the stem with ease.During the (b)(6) 2022, revision surgery, surgeon removed the wright medical profemur® hip system, including the profemur® plus cocr modular neck and profemur® tl stem, as well as the conserve® femoral head and conserve® neck sleeve, all of which had been implanted in patient's right hip in 2012.These parts were then replaced with a femoral stem, femoral head, cup and liner, all manufactured by smith & nephew.Patient still experiences residual symptoms following the (b)(6) 2022, right hip revision surgery.Patient suffered a subsequent post-revision dislocation of his right hip on or about (b)(6) 2022.Surgeon has also determined that patient suffered significant posterior hip muscle loss, bone loss, loss of abductor function, and impairment of patient's gait and right hip stability.It was indicated on litigation file that patient's profemur® plus cocr modular neck suffered "micromotion, fretting corrosion, galvanic corrosion, and fracture".Also, it was indicated that due to micromotion and fretting corrosion of the oblong taper of the profemur® plus cocr modular neck where it seated in the pocket of the stem, the profemur® plus cocr modular neck implanted in patient's right hip corroded, causing physical injury to patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROFEMUR® TL STEM SZ 4
Common Device Name
HIP COMPONENT
Manufacturer (Section D)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC.
5677 airline rd.
arlington TN 38002
Manufacturer Contact
5677 airline road
arlington, TN 38002
9018674771
MDR Report Key15419082
Report Number3010536692-2022-00332
Device Sequence Number19488105
Product Code JDI
UDI-Device IdentifierM684PRTL00241
UDI-PublicM684PRTL00241
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K060358
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberPRTL0024
Device Catalogue NumberPRTL0024
Device Lot Number1450581
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/23/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-