• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT PS POR FEM PS POR LEFT SZ 3.5; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT PS POR FEM PS POR LEFT SZ 3.5; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED Back to Search Results
Model Number TRULIANT TIB IMP PS INSERT SZ 3.5 9MM
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Unspecified Tissue Injury (4559); Aseptic loosening (4883)
Date of Event 08/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, the patient had an initial left tka on (b)(6) 2021.The patient complained of sever pain, a possible loose femur, and was revised to a competitor's product.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.
 
Additional Manufacturer Narrative
Pending investigation.Concomitant medical products: 6739119, 02-020-12-0235 - truliant ps por fem ps por left sz 3.5; 6030524, 02-022-55-3525 - truliant por tib tray size 3.5f/25t; ;6982635, 200-07-32 - advanced patella 32mm 3 peg implant.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of an insufficient bond between the femoral component and the bone, which led to aseptic (non-infected) femoral loosening, osteolysis, and pain.However this cannot be confirmed as the devices were not returned for evaluation and pre-revision radiographs were unable to be obtained.H7: z-0023-2022.
 
Event or Problem Description
As reported, the patient had an initial left (b)(6).The patient complained of sever pain, a possible loose femur, and was revised to a competitor's product.The patient was last known to be in stable condition following the event.No other patient information/medical history reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b5, d8, f10 (remove data from initial), g1, h6: clinical code, component code, type of investigation, investigation findings, investigation conclusions.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT PS POR FEM PS POR LEFT SZ 3.5
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15419116
Report Number1038671-2022-01112
Device Sequence Number14789085
Product Code MBH
UDI-Device Identifier10885862314796
UDI-Public10885862314796
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K181794
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/17/2026
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTRULIANT TIB IMP PS INSERT SZ 3.5 9MM
Device Catalogue Number02-020-12-0235
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer03/14/2023
02/05/2026
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date04/01/2023
02/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-