• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION XXL VASCULAR; DILATOR, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION XXL VASCULAR; DILATOR, ESOPHAGEAL Back to Search Results
Model Number 23375
Medical Device Problem Codes Inflation Problem (1310); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The patient underwent percutaneous coronary intervention (pci).The target lesion was located in the iliac vein.Vascular access was obtained via the left femoral vein and a 10f sheath was inserted.After advancing a 260 cm guidewire to the inferior cavity, a conical artery catheter was exchanged with a stiffened guidewire.A 4-6/5.8/75 xxl vascular balloon catheter was advanced for dilatation.During digital subtraction angiography (dsa), it was found that the rated burst pressure of the balloon was 8 atmospheres.However, during inflation, the pressure could not go higher than 4 atmospheres.Consequently, the balloon was not fully inflated.Upon withdrawal, it was found that the balloon was leaking as the contrast agent had mixed with blood.A syringe was used to inject a liquid into the balloon outside the patient and a pinhole was noted.The device was replaced with a 16x4 xxl vascular balloon catheter and then a 16x90 boston scientific stent was implanted to complete the procedure.No patient complications were reported and the patient status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: a xxl device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 8mm in length and extended from a position 20mm distal of the proximal markerband to 29mm distal of the proximal markerband.The rated burst pressure for this device is 8 atmospheres as per xxl specification.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.
 
Event or Problem Description
It was reported that balloon rupture occurred.The patient underwent percutaneous coronary intervention (pci).The target lesion was located in the iliac vein.Vascular access was obtained via the left femoral vein and a 10f sheath was inserted.After advancing a 260 cm guidewire to the inferior cavity, a conical artery catheter was exchanged with a stiffened guidewire.A 4-6/5.8/75 xxl vascular balloon catheter was advanced for dilatation.During digital subtraction angiography (dsa), it was found that the rated burst pressure of the balloon was 8 atmospheres.However, during inflation, the pressure could not go higher than 4 atmospheres.Consequently, the balloon was not fully inflated.Upon withdrawal, it was found that the balloon was leaking as the contrast agent had mixed with blood.A syringe was used to inject a liquid into the balloon outside the patient and a pinhole was noted.The device was replaced with a 16x4 xxl vascular balloon catheter and then a 16x90 boston scientific stent was implanted to complete the procedure.No patient complications were reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XXL VASCULAR
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15419172
Report Number2124215-2022-33704
Device Sequence Number12416827
Product Code LIT
UDI-Device Identifier08714729190486
UDI-Public08714729190486
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/26/2025
Device Model Number23375
Device Catalogue Number23375
Device Lot Number0028949226
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/25/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight60 KG
-
-