| Model Number |
AR40E |
| Medical Device Problem Codes |
Material Fragmentation (1261); Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the leading haptic of the intraocular lens (iol) was detached inadvertently when about half of the lens was already inside the patient's right eye.The incision was enlarged.The lens and the detached haptic were removed from the eye.Vitrectomy was performed and sutures were placed.There was a delay of approximately 45 minutes and medication was prescribed.The patient was left aphakic as there was no suitable lens (a 3 piece iol) available during the surgery to be used as the replacement iol.The patient remained aphakic for 7 weeks ((b)(6) 2022 to (b)(6) 2022) and had regular follow-up consultations during this time.Eventually, the patient was implanted with a 3 piece lens of the same power as the original lens.The original lens was discarded.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable, as the lens was not implanted.If explanted, give date: not applicable, as the lens was not implanted.Telephone number: (b)(6).The device was not returned for evaluation as the lens was discarded.Therefore, a failure analysis of the complaint device cannot be completed.A review of the photo, device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Health effect - impact code: (b)(4).Health effect - clinical code: (b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information from the facility was received and revealed following information: section d10: concomitant medical products: emerald cartridge, emeraldt handpiece.Device evaluation: product evaluation was not performed because the product has not been received.Visual inspection of the customer provided photo revealed the complaint lens inside of the lens case and coated in viscoelastic residue.One haptic can also be observed to be detached from the lens and bent.No further issues could be observed with the photo provided.The complaint issue reported could not be confirmed and no product deficiency could be identified.Conclusion: the complaint issue "delivery issue" was not confirmed.The complaint issue "haptic detached" was observed during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|