INTEGRA LIFESCIENCES SWITZERLAND SAR PROG VALVE INLINE W SG; CHPV WITH SG Back to Search Results
Model Number 823162
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Fever (1858); Vomiting (2144)
Date of Event 08/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A physician reported a hakim valve (id 823162) was implanted in a 2 month old patient via unknown shunt and date.On (b)(6) 2022, the patient had a subcutaneous retention, large fontanel swelling, fever, and vomiting.On (b)(6) 2022, shunt imaging was performed and obstruction was confirmed.Therefore, the valve was removed and replaced on (b)(6) 2022.The patient is in follow up.
 
Additional Manufacturer Narrative
The hakim valve (id 823162) was returned for evaluation.Device history record (dhr) - the product code 82-3162 with lot 6317112, conformed to the specifications when released to stock.Failure analysis - the valve was visually inspected; a needle hole in the needle chamber was noted.The position of the cam when valve was received was 70mmh2o.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problems with the valve at the time of investigation.The possible root cause for the issue reported by the customer, could be due to biological debris and protein build up interfering with the valve mechanism at the time of investigation no occlusion issues were noted with the valve.No corrective action is required given that no manufacturing, workmanship, material deficiency or adverse trend was identified.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROG VALVE INLINE W SG
Common Device Name
CHPV WITH SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15419402
Report Number3013886523-2022-00424
Device Sequence Number6526259
Product Code JXG
UDI-Device Identifier10381780519232
UDI-Public(01)10381780519232(17)270331(10)6317112
Combination Product (Y/N)N
PMA/510(K) Number
K041296
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number823162
Device Catalogue Number823162
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age2 MO
Patient SexUnknown
-
-