| Model Number |
3125-1200S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/30/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.Device disposition is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient had a left hip subtrochonteric fracture.Patient also had a previous revision total knee with a long stem on the femoral component.On (b)(6) 2022 the patient went into office and it was noted on x rays that the fracture had fallen slightly into varus.Still intact.A broken nail can be seen on the picture.He was to be on 50% weight baring status.Patient claimed he could not maintain his 50% weight baring status due to living alone.He had a walker but was not using it.He has a history of non compliance post op.He had an additional surgery performed at another facility.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed from the x-rays that were provided and matches the alleged failure mode.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Multiple x-rays were provided which were presented to our health care professional for evaluation, and a question about the most likely reason for the breakage was asked, to which the medical expert replied: ¿there are some patient related factors: severe overweight with a bmi of 35,8.The co-morbidities indicate reduced compliance (substance abuse, schizophrenia).The fracture pattern with a pretty straight subtrochanteric location is also unfavorable.Surgery related factors: reduction of the fracture is (probably also due to the overweight) pretty poor.There is quite a rotational and axial dislocation after insertion of the nail.The nail lies eccentric within the neck of the long neck/head fragment.This position prevents good healing due to movement of the fragments during the mobilization in the hip even without weight bearing.The points seem to be the most relevant one.Others could be detected by inspection of the nail.However, these are enough points to explain the failure.¿ [original statement(s)].More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However, based on the information provided and the medical expert¿s opinion the issue seems to be a patient-related factor.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient had a left hip subtrochonteric fracture.Patient also had a previous revision total knee with a long stem on the femoral component.On (b)(6) 2022 the patient went into office and it was noted on x rays that the fracture had fallen slightly into varus.Still intact.A broken nail can be seen on the picture.He was to be on 50% weight baring status.Patient claimed he could not maintain his 50% weight baring status due to living alone.He had a walker but was not using it.He has a history of non compliance post op.He had an additional surgery performed at another facility.
|
| |
|