• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SMALL BATTERY DRIVE CASING F/ 14.4 VOLT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SMALL BATTERY DRIVE CASING F/ 14.4 VOLT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.132
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Premature Separation (4045)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported from singapore that during service and evaluation, it was determined that the small battery drive device had unintended/unexpected disengagement and component damage.It was also observed that the device casing was unable to attach to handpiece device as the pins were broken.It was further determined that the device failed pretests for function of battery casing cover and connection battery casing to handpiece.It was noted in the service order that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE CASING F/ 14.4 VOLT
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key15419605
Report Number8030965-2022-06809
Device Sequence Number2545416
Product Code HWE
UDI-Device Identifier07611819418301
UDI-Public07611819418301
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number532.132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/22/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2022
Initial Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-