Model Number 325110
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the pressure pump in the eagle inflation device leaked water when in use by the clinician.It was mentioned that a new pressure pump was used.
 
Event or Problem Description
It was reported that the pressure pump in the eagle inflation device leaked water when in use by the clinician.It was mentioned that a new pressure pump was used.As per follow up information received via ibc on 15sep2022, it was stated that was no impact to the patient.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned and further evaluation was not conclusive.Although a specific cause cannot be determined, based on the reported event a potential root cause could be that the device was damaged in transit.It was unknown if the device had met all relevant specifications or resulted in the reported event.No manufacturing issues or non-conformances were noted during review of the dhr that could have caused or contributed to the reported event.The instructions for use were found adequate and states the following: "before use, inspect the device to verify that no damage has occurred during shipping and handling." "attaching the inflation device to the balloon dilation catheter: 1.Prepare and test the balloon dilation catheter according to the manufacturer¿s directions for use.2.If a separate syringe was used to prepare the balloon catheter, remove it.Create a fluid-fluid connection between the balloon and the connecting tube (male rotating adapter) of the inflation device by placing a drop of contrast solution from the syringe into each hub.3.Hand tighten the hubs securely." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® EAGLE® INFLATION DEVICE
Common Device Name
EAGLE INFLATION DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15419689
Report Number1018233-2022-07132
Device Sequence Number12393614
Product Code KOE
UDI-Device Identifier00801741067242
UDI-Public(01)00801741067242
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K962611
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model Number325110
Device Catalogue Number325110
Device Lot NumberBMFQAT02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/05/2022
Supplement Date Received by Manufacturer02/10/2023
Initial Report FDA Received Date09/14/2022
Supplement Report FDA Received Date02/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-