BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 87047
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to boston scientific for analysis.Visual inspection revealed the irrigation extension tubing was found partially detached/separated from the luer fitting at the adhesive joint.Functional test revealed no abnormal resistance was felt when actuating the steering mechanism.The continuity test was performed and no problems were detected.Analysis of returned product revealed that the unit showed no evidence or defects that could have contributed to the signal issue, therefore, not confirming the reported complaint.
 
Event or Problem Description
Reportable base on analysis completed on (b)(6) 2022.Prior to a radiofrequency ablation procedure a intellanav mifi open-irrigated catheter was selected for use.The signal was not stable.An exchange of the catheter resolve the problem.The procedure was completed without any patient complications.However, analysis of the returned device revealed that the irrigation extension tubing was found partially detached/separated from the luer fitting at the adhesive joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key15420435
Report Number2134265-2022-08781
Device Sequence Number2621349
Product Code OAD
UDI-Device Identifier08714729938378
UDI-Public08714729938378
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P150005/S017
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0027590966
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/18/2022
Initial Report FDA Received Date09/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight52 KG
-
-