| Model Number |
DTBA1QQ |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Electromagnetic Interference (1194); Over-Sensing (1438)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Dyspnea (1816)
|
| Date of Event |
03/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 0296 lead implanted: (b)(6) 2013.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) exhibited electromagnetic interference (emi) oversensing while the patient was working on copper sink plumbing near electrical wiring.The patient reported feeling "weird" with palpitations and light headedness.The patient's medication was adjusted and the crt-d was reprogrammed.The patient is a participant in a clinical study.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|