| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of reft1133 showed four other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported that at the end of the device placement procedure, when it was being tested for flow and reflow, 0.9% saline solution was infused using a 10ml syringe, it was observed that the catheter was leaking due to a hole in it.It was stated the picc with leakage was removed and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that at the end of the device placement procedure, when it was being tested for flow and reflow, 0.9% saline solution was infused using a 10ml syringe, it was observed that the catheter was leaking due to a hole in it.It was stated the picc with leakage was removed and replaced.Additional information received: reported hole was at the 2cm mark.
|
| |
|