| Model Number |
TRM-310 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/15/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported a left side exchange from textured to smooth breast implants due to the patient¿s concern with the product and rupture.This record is for the left side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: exchange from textured to smooth breast implants due to the patient¿s concern with the product and rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported a left side exchange from textured to smooth breast implants due to the patient¿s concern with the product and rupture.This record is for the left side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed, broken device on the posterior side assessed as surgical impact opening (shell thickness within specification).Opening device on the posterior side assessed as surgical impact opening (shell thickness within specification).Anxiety¿product/procedure: unable to observe since it is a medical event and is not related to the device.Additional observations: nick is observed assessed as non-penetrating nick.No further actions are required as the device was implanted.
|
| |
|